AMI Codes-II O3 is a proven, low-maintenance monitor for the measurement of trace concentrations down to 1 ppb of ozone in water for pharmaceutical purposes. Using the robust colorimetric standard measurement method (DPD) according to DIN 38404-3, it is ideally suited to reliably prove the continuous protection of water storage systems against microbial growth as well as the absence of ozone before the water is used in a pharmaceutical production process.
Features
Your benefits
Keywords: Ozone analyzer, ozone sanitization, water for pharmaceutical purposes, purified water (PW), highly purified water (HPW), water for injection (WFI), water distribution and storage, on-line measurement
Ozone measurement | |
Measuring range | 0 to 500 ppb |
Accuracy | ±1 ppb or 5% (whichever is greater) |
Limit of detection | 1 ppb |
Cycle time | 5 ... 10 min. (default 6') |
Transmitter Specifications |
|
Power Supply | 100 ... 240 VAC, 50/60Hz or 24 VDC |
Power consumption | max. 30 VA |
Sample conditions |
|
Water consumption | min. 10 l/h |
Water inlet pressure | 0.15 ... 2 bar |
Sample temperature | 5 ... 50 °C |
Panel |
|
Dimension | 400 x 850 x 200 mm |
Material | stainless steel |
Weight | 14.0 kg |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有