Colorimetric process analyzer (DPD-method) for measurement and dosing control of disinfectants such as free chlorine, hypochlorus acid, chlorine dioxide, ozone, bromine, iodine and monochloramine.
Options:
Disinfectant measurement | Measuring range [ppm] | Accuracy | |
Ozone | 0.05 | 1.00 | ±0.01 ppm |
HOCL, free chlorine, monochloramine | 0.00 | 1.00 | ±0.01 ppm |
1.00 | 3.00 | ±0.06 ppm | |
3.00 | 5.00 | ±0.2 ppm | |
Chlorine dioxide, iodine, bromine | 0.00 | 2.00 | ±0.02 ppm |
2.00 | 6.00 | ±0.12 ppm |
Response time | 90% of change of excessive Cl in 60 seconds after sample entered flow cell. |
Cycle time free chlorine | 1 ... 12 min. |
pH measuring range | 2 ... 12 |
Temperature measuring range | -30 ... 100 °C |
Transmitter Specifications |
|
Power Supply | 100 ... 240 VAC, 50/60Hz or 24 VDC |
Power consumption | max. 30 VA |
Sample conditions |
|
Water consumption | min. 10 l/h |
Water inlet pressure | 0.15 ... 2 bar |
Sample temperature | 5 ... 50 °C |
Panel |
|
Dimension | 400 x 850 x 200 mm |
Material | PVC |
Weight |
9.0 kg |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有