Description
Hand-held instrument for the determination of disinfectants, dissolved iron, dissolved aluminium, pH-value and ORP/redox-potential.
-
Photo- and pH-combimeter with state of the art microelectronics and clearly arranged display.
-
Well-devised analytical set in portable case.
-
Choice between routine and expert mode.
-
Determination of pH-value and ORP/redoxpotential by maintenance-free electrodes.
-
Determination of disinfectants with Oxycon reagents following the DPD method
-
Determination of free and bound chlorine, chlorine dioxide, ozone, iodine, bromine, cyanuric acid, dissolved iron, dissolved aluminium, pH-value and ORP/redoxpotential.
Options:
-
Reagent set's.
-
pH measurement (electrode with cable, BNC connector).
-
ORP/redox measurement (electrode with cable, BNC connector.
Specifications
Scope of supply:
Combimeter in carrying case (38x27x10 cm) with electronic thermometer, 4 glass cuvettes, 1 cuvette lid, 2 glass rods, 1 ball pipette, 3 sample flasks, 1 waste flask, 1 note pad, 1 waterproof pen, 4 batteries.
Instrument:
-
Combined photo- and pH-meter, microprocessor controlled
-
Digital display
-
Battery powered with economical power circuit for about 2000 determinations
Dimensions, weight:
-
Instrument: 10 x 20 x 4 cm; 450 g
-
Carrying case: 38 x 27 x 10 cm; 2.900 kg (complete)
Specifications pH:
-
Maintenance-free pH electrode with cable and BNCconnector
-
Measuring range: 0 - 14 pH
-
Resolution: 0.01 pH
-
Accuracy: ± 0.01 pH
-
Automatic temperature compensation
-
Automatic buffer recognition
-
ISO/DIN or technical buffer
Specifications temperature:
-
Electronic thermometer
-
Measuring range: - 50 to + 170 °C
-
Resolution: 1 °C
-
Accuracy: ± 1 °C
-
Probe: InAisi316, length 130 mm, ∅ 3 mm
-
Size: 142 x 29 x 15 mm
-
Battery: 4 x 1.4 V
-
Battery life: approx. 1000 hours
Specifications ORP/redox (optional):
-
Maintenance-free ORP/redox electrode with cable and BNC-connector
-
Measuring range: - 400 mV to + 1500 mV
-
Resolution: 1 mV
-
Accuracy: ± 0.5 mV
-
Automatic temperature compensation
-
Automatic buffer recognition
-
ISO/DIN or technical buffer
Photometric specifications:
Parameter
|
Measuring range mg/l |
|
low |
medium |
high |
Chlorine |
0.00 |
- 2.49 |
- 4.9 |
- 10 |
Chlorine-dioxide |
0.00 |
- 4.99 |
- 9.9 |
- 20 |
Ozone |
0.000 |
- 0.499 |
- 0.99 |
- 2.5 |
Iodine |
0.00 |
- 9.99 |
- 19.9 |
- 35 |
Bromine |
0.00 |
- 4.99 |
- 9.9 |
- 20 |
Accuracy: |
|
± 1% |
± 2.5% |
± 5 % |
|
|
of measuring range |
Parameter
|
Measuring range
|
Accuracy
(of measuring range) |
pH-value |
6.50 |
- 8.00 pH |
± 0.1 pH |
Cyanuric acid |
0 |
- 100 mg/l |
± 10% |
Iron |
0.00 |
- 2.5 mg/l |
± 1% |
Aluminium |
0 |
- 0.8 mg/l |
± 5% |
-
Repeatability better than accuracy
-
Automatic range switching