Sensor for the measurement of specific conductivity.
For applications in surface water, potable water and cooling water.
For the use together with SWAN conductivity transmitters AMI Solicon4 or AMU Solicon4.
Recommended conductivity range | 0.1 µS/cm ... 100 mS/cm |
Accuracy | ± 1% or ± 0.2 µS/cm; whichever is greater |
Cell constant k | 0.45 cm-1 |
Temperature sensor type | Pt1000, DIN class A |
Operating conditions: |
|
- max. temperature | 120°C at 6.5 bar |
- max. pressure | 12 bar at 20°C |
- sterilizable at | 120°C / 5 bar / 30min. |
Materials: |
|
- Body | PPSU |
- Shaft | stainless steel SS316L |
- Current electrodes | SS 316L |
- Isolation | PEEK |
- Voltage electrodes | Platinum |
- Cover | PVDF |
Electrical connection |
Plug M16 male (IP67) |
Process connection | G1" thread |
Free space around sensor tip | 20 mm |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有