Dual channel monitor for the continuous measurement of pH and Redox (ORP) potential in pool water.
Complete system mounted on PVC panel including:
For use with (order seperately):
Optional accesory:
pH Measurement with Swansensor pH Standard | |
Measuring range | pH 1 to 13 |
Resolution | 0.01 pH |
Reference temperature | 25° C |
Automatic temperature compensation according to Nernst. | |
ORP (Redox) measurement with Swansensor Redox Standard | |
Measuring range | -400 to 1'200 mV |
Resolution | 1 mV |
Temperature measurement with Swansensor Temperature Standard | |
Measuring range | -30 to 130°C |
Resolution | 0.1° C |
Power Supply | |
Voltage |
100-240 VAC (+- 10%), 50/60 Hz (+- 5%) or |
Sample conditions | |
Flow rate | min. 30 l/h |
Temperature | up to 50° C |
Inlet-, Outlet pressure | 0.2 to 2 bar |
Dimensions | |
Panel | 300 x 530 x 150 mm |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有