Monitor for the continuous measurement of dissolved hydrogen in water steam cycles. Integrated auto verification with faraday electrode setup.
Application
Parameters such as pH, oxygen, iron, hydrogen and others are directly related to corrosion control and frequently used for monitoring corrosion.
Hydrogen monitoring is becoming important in corrosion monitoring for water steam cycle. Corrosion occurs in the presence of water contact with iron at high temperature. According to the Schikorr reaction, hydrogen is released as a by-product during the formation of the magnetite and hematite layer in the power plant. Therefore its concentration is a direct measure of the rate of corrosion.
Dissolved Hydrogen measurement | |
Measuring range | 0,1 to 800 ppb or 0 to 50% saturation |
Resolution |
0.01 ppb (up to 9.99 ppb) 0.1 ppb (from 10.0 to 99.9 ppb) 1 ppb (from 100 to 800 ppb) |
Accuracy |
+- 5% of reading or +- 0,5 ppb (whichever is greater) |
Repeatability |
+- 1% of reading or +- 0,5 ppb (whichever is greater) |
Temperature measurement | |
Measuring range | -30 to 130°C |
Resolution | 0.1° C |
Power Supply | |
Voltage |
100-240 VAC (+- 10%), 50/60 Hz (+- 5%) or 24 VDC (+- 10%) |
Sample conditions | |
Flow rate | 6 to 20 l/h |
Temperature | up to 45° C |
Inlet pressure | up to 1 bar |
Outlet pressure | pressure free |
Suspended solids | less than 10 ppm |
Dimensions | |
Panel | 280 x 850 x 150 mm |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有